HILL AIR FLEX
UNPLUG TABLE BEFORE STARTING ANY REPAIRS
PROBLEM: DROP PIN OUT OF CHANNEL
1.
Unplug
the table
2.
Locate
the coil that needs replacing(if you have a coil with wires coming directly out
of it, locate where the wires are already connected and cut the
colored wire (not the white one) close to the coil
(photo 1).
If your coil already has wires
with connectors on them, then
proceed to step 4.
3.
Take the
wire you cut and strip
back the outer layer approx. 3/8”(photo 2)

and put on the ¼”
connector with a wire crimper
(photo 3)

4.
Remove
the nut on top of the old coil with a 9/16” wrench(photo 4)

and take the old
coil off of the
valve stem. If the old coil has a metal bracket on
it, do NOT use the metal bracket with the new coil. Pull the wires off
of the coil (be careful to pull
the wire connectors and not
the wires.
5.
Place the
new coil on the valve stem and replace the 9/16” nut.
6.
Push the
wires with the wire
connectors onto the new coil
|